The new generation cooling/heating forcing system is designed as laboratory equipment to validate Si components within a temperature range of -60 ºC up to +150 ºC, with a cooling/heating envelope of 350W@-40ºC, 1200W@0ºC, and 1200W@140 ºC. The Samson+ system supports rack dimensions of 50 x 25 x 35 cm and requires a standard 208–230V 50Hz, 230V 60Hz, or 3 phase 440V 50/60Hz power outlet. Samson liquid cooling requires an additional chiller with a flow rate range of 2.0–3.5 liters per minute @15 ºC (+/- 2 ºC). Total cooling system efficiency is approximately 55%.
| Temperature range | -60°C to 150°C |
|---|---|
| Cooling power | 350W @ -40/0°, 500W @ -24/0°, 1200W @ 0°C, 1200W @ 140° |
| Head size | 78 × 90 mm |
| Temperature change rate | ~ (hot/cold): 3°C per second |
| Condensation prevention | Air Purge & Heaters |
| Umbilical hose length Remote | 2.3 meters (between control box & thermal head) |
| Remote communication | I^2C / Modbus TCP/UDP Protocols |
| Refrigeration | R452a – complies with CA state GWP regulation (2020) |